DC Current Output
To set dc a current output at the Calibrator AUX or
20A terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure dc current.
4. Enter the desired current output (up to 6
digits).
5. Press I to select polarity.
6. Press a multiplier key, if necessary.
7. Press A.
8. Control Display shows entered value. For
example, 234.567 mA.
oe32f.eps
9. Press E.
oe33f.eps
10. Press O to activate the Calibrator output.
OUTPUT selects the AUX or 20A terminals.
Outputs 3A or above are always on the 20A
terminals.
Press the auto/locked softkey to toggle between
autorange mode or lock the Calibrator in the
present range.