Multiplying and Dividing
Press X to multiply the output by 10. Press
D to divide the output by 10. If multiplied output
exceeds 33 V, the Calibrator is placed in STBY
(Standby).
Setting Voltage and Current Limits
Output limits help prevent accidental damage to
UUTs from overcurrent or overvoltage. Selections
are saved in nonvolatile memory. Voltage limits
are expressed in rms.
To set voltage and current limits.
1. Press R to clear the Calibrator output.
2. Press S. Press INSTMT SETUP to open
setup submenus.
3. Press OUTPUT SETUP to open output setup
submenus.
4. Press SET LIMITS to open set limits menu.
oe74f.eps
5. To limit voltage(applies to both dc and ac
voltages), press the softkey under VOLTAGE.
oe75f.eps