6270A
Operators Manual
10
Product Placement
Warnings
To prevent possible electrical shock, fire, or personal injury, do
not restrict access to the Product mains power cord. The mains
power cord is the mains disconnecting device. If access to the
power cord is inhibited by rack mounting, a properly-rated
accessible mains disconnecting switch must be provided within
reach as part of the installation.
To prevent possible personal injury:
Use good lifting practices when lifting or moving the
Product. The Product is an unbalanced load and can weigh
as much as 20 kg (44 lb).
Use two handles to lift and carry the Product.
Use the Product on a bench or in a standard 19 inch equipment rack. Purchase a
rack mount kit to install the Product into an equipment rack. For bench-top use,
install the Product on a flat, stable surface at a convenient height. The front feet
can be extended to incline the Product for easier viewing.
Minimize the distance between the Product and the device or system under test
to enhance control performance and reduce pressure settling times.
For installation, the Product requires:
An electrical power source of 100 V ac to 240 V ac, 47 Hz to 63 Hz.
A continuous, regulated pressure supply of clean, dry, non-corrosive gas at
the Product’s maximum control pressure +10 % or 70 kPa (10 psi), whichever
is greater, to be connected to the Product SUPPLY port.
A vacuum source of 7 kPa (1 psi) absolute and with a displacement of at
least 90 L/m (3 cfm) if controlling pressures <20 kPa (3 psi) gauge.