Pressure Controller/Calibrator
Troubleshooting
71
Remote Interface
The Remote Interface Diagnostics menu provides information on remote
communication by monitoring the input and output communication. The menu
contains independent diagnostics for RS-232, USB, Ethernet, GPIB, and the
System Bus communications.
RS-232 – Shows the Receive and Transmit data for this interface.
USB – Shows the Receive and Transmit data for this interface.
Ethernet – Shows the Receive and Transmit data for this interface.
GPIB – Shows the Receive and Transmit data for this interface.
System Bus – Shows the subsequent data for this interface.
Receive
Transmit
Fault
Troubleshooting
Table 17 explains minor troubleshooting issues. For issues outside of the scope of this
section, the Product may require service. See How to Contact Fluke Calibration.
Table 17. Troubleshooting
Problem Probable Cause Action
Electrical Problems
Product
does not
turn on
Not plugged in Verify Product is plugged in and power is available.
Power not available Check the fuse per instruction in this manual.
Fuse blown
Verify the main power switch is ON.
Check the fan. If the fan is on or solenoid click is heard, see
the “Display does not turn on” problem.
If the fan is not on, have the power supply serviced. Verify
internal power connections.
Display
does not
turn on
Chassis power issue Verify that power is supplied, see above.
Screen saver is
activated
Verify power to front panel. and should be lit.
Power plug to front
panel is loose
Verify the is in the ON state.
Display panel failure
If is yellow, then push it to turn on display.
Open front panel and verify the power plug is connected to
the front panel.