Pressure Controller/Calibrator
CPS Use
53
P3000
Note
The thread on the test port, and the lower part of the gauge adapters
is LEFT-HANDED. The following procedure details the correct
method for mounting devices using these adapters:
1. Screw the appropriate gauge adapter fully on to the UUT, see Figure 10.
hwr029,eps
Figure 10. Screw on Gauge Adapter
2. Screw assembly down COUNTER-CLOCKWISE on to test port, see
Figure 11.