6270A
Operators Manual
56
Table 15. Test Port Inserts - Parts List
Item Description Part
1 O-ring 3865163
2 Test Port Insert 3919892
3 O-ring 3865195
4 Test Port 4542465
To calibrate panel-mounted gauges with pressure connections in the rear, use an
Angle Adapter such as the Fluke P5543. The Product operates the CPS without
any further interaction required by the operator.
Disconnect the CPS
To disconnect the CPS, ensure that the system is vented then disable the CPS in
the settings. It is important that the CPS is disabled in the firmware if it is
physically disconnected from the system. Failure to do so could result in the
system not being able to vent pressure. Once disabled in the firmware, the CPS
can be disconnected both electrically and pneumatically.
External Driver Communication
The Product has four external 24 V dc powered drivers. These drivers support
several optional accessories and can support the driving of solenoids for
customized solutions. The driver status can be set with the display and remote
interfaces.
Each driver is a low-side switched 24 V dc component. Each Channel includes a
PTC-type resettable fuse.
Maximum continuous power per channel is 10 W and total continuous power
maximum for all 4 channels combined is 24 W. See Figure 15.
Drivers
huo030.eps
Figure 15. Drivers