71X Series
Calibration Manual
10
Accuracy (˚C) *
Input
RTD
Type
Range ˚C
4-Wire 2-Wire & 3-Wire
Source
Allowable
Excitation (mA)
-200.0 to 100.0 0.20 0.25 0.2
100.0 to 300.0 0.30 0.40 0.3
Pt 1000 385
300.0 to 630.0 0.40 0.52 0.4
0.05 to 0.4
-200.0 to 100.0 0.20 0.28 0.2
100.0 to 300.0 0.30 0.40 0.3
Pt 100 392
(3926)
300.0 TO 630.0 0.40 0.52 0.4
0.1 to 3.0
-200.0 to 100.0 0.20 0.28 0.2
100.0 to 300.0 0.30 0.40 0.3
Pt 100 JIS
3916
300.0 to 630.0 0.40 0.52 0.4
0.1 to 3.0
Addresses pulsed transmitters and PLCs with pulses as short as 5 ms.
Excitation current from 712: 0.2 mA.
Maximum input voltage: 30 V
*2-wire: Does not include lead resistance
3-wire: Assumes matched leads
Table 9. Ohms Measurement Specifications
Accuracy *
Ohms Range
4-Wire 2- and 3-wire
0 to 400 0.025 % ±0.05 0.025 % ±0.1
400 to 4000 0.025 % ±0.05 0.025 % ±0.55
Excitation current: 0.2 mA.
Maximum input voltage: 30 V
*2-wire: Does not include lead resistance
3-wire: Assumes matched leads
Table 10. Ohms Source Specifications
Ohms Range
Excitation Current from
Measurement Device
Accuracy
5 to 400 0.1 to 0.5 mA 0.025 % ±0.1
5 to 400 0.5 to 3.0 mA 0.025 % ±0.05
400 to 1500 0.05 to 0.8 mA 0.025 % ±0.5
1500 to 4000 0.05 to 0.4 mA 0.025 % ±0.5
Resolution
RTD: 0.1 °C
Ohms: 0.1
Rev 1.3 or Later: < 400 0.01
> 400 0.1