Process Calibrators
Specifications
11
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges –10 °C to 18 °C and 28 °C to
55 °C. Ohms ranges are 400 Ω, 1.5 kΩ, and 4.0 kΩ.
713 Specifications
Table 11. 713 30G Pressure Input
Firmware Version* Range Accuracy
V1.2 or earlier 0 to 30 psi (206.85 kPa) ± 0.05 % of range
V1.3 or later -12 psi (82.74 kPa) to 30 psi (206.85 kPa) ± 0.05 % of range
Maximum nondestructive pressure: 3X top of range (90 psi, 620 kPa, 6.2 bar)
Temperature coefficient: 0.01 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55°C
* To display the firmware version, start with the 713 off, press and hold M, then press the power button.
Table 12. 713 100G Pressure Input
Firmware Version* Range Accuracy
V1.2 or earlier 0 to 100 psi (689.5 kPa) ± 0.05 % of range
V1.3 or later -12 psi (-82.8 kPa) to 100 psi (689.5 kPa) ± 0.05 % of range
Maximum nondestructive pressure: 2X top of range (200 psi, 1380 kPa, 13.8 bar)
Temperature coefficient: 0.01% of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C
* To display the firmware version, start with the 713 off, press and hold M, then press the power button.