7381
User’s Guide
44
Table 5. Interface Command Summary
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint] s set: 9999.99 {C or F} set: 150.00 C
Set current set-point to n
s[etpoint]=n
t[emperature]=n
s=150
t=150
Instrument Range
-40°C to 150°C
Read vernier
v[ernier] v v: 9.99999 v: 0.00000
Set vernier to n
v[ernier]=n v=.00018
Depends on
Configuration
Read scan function
sc[an] sc scan: {ON or OFF} scan: ON
Set scan function:
sc[an]=on/of[f] ON or OFF
Turn scan function on
sc[an]=on sc=on
Turn scan function off
sc[an]=of[f] sc=of
Read scan rate
sr[ate] sr
srat: 9.999
{C or F}/min
srat: 0.010
C/min
Set scan rate to n degrees per
minute
sr[ate]=n sr=5
0.001 to 5.000°C/min
0.002 to 9.000°F/min
Read temperature
t[emperature] t t: 9999.99 {C or F} t: 55.69 C
Read temperature units
u[nits] u u: x u: C
Set temperature units:
u[nits]=c/f C or F
Set temperature units to Celsius
u[nits]=c u=c
Set temperature units to Fahrenheit
u[nits]=f u=f
Ramp and Soak Menu
Read number of programmable
set-points
pn pn pn: 9 pn: 2
Set number of programmable
set-points to
n
pn=n pn=4 2 to 8
Read programmable set-point
number n
psn ps3 psn: 9999.99 {C or F} ps1: 50.00 C
Set programmable set-point
number n to n
psn=n ps3=50
1 to 8, Instrument
Range –40 °C to
150 °C
Read program set-point soak time
pt pt ti: 999 ti: 5
Set program set-point soak time to
n minutes
pt=n pt=5 0 to 500
Read program control mode
pc pc prog: {OFF or ON} prog: OFF
Set program control mode:
pc=g[o]/s[top]/c[ont]
GO or STOP or
CONT
Start program pc=g[o] pc=g
Stop program pc=s[top] pc=s
Continue program pc=c[ont] pc=c
Read program function pf pf pf: 9 pf: 3
Set program function pf=n pf=2 1 to 4