Calibration Bath
Digital Communication Interface
45
Table 5. Interface Command Summary (cont.)
Command Description Command Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Secondary Menu
Read proportional band setting pr[op-band] pr pb: 9.999 pb: 0.315
Set proportional band to n pr[op-band]=n pr=0.326
Depends on
Configuration
Read cutout setting c[utout] c cu: 999 {x},{xxx} cu: 120 C, in
Set cutout setting: c[utout]=n/r[eset]
Set cutout to n degrees c[utout]=n c=120
Temperature
Range
Reset cutout now c[utout]=r[eset] c=r
Read heater power
(duty cycle)
po[wer] po po: 9999 po: 1
Configuration Menu
Probe Menu
Read R0 calibration parameter r[0] r r0: 999.999 r0: 100.578
Set R0 calibration parameter to
n
r[0]=n r=100.324 98.0 to 104.999
Read ALPHA calibration
parameter
al[pha] al al: 9.9999999 al: 0.0038573
Set ALPHA calibration parameter
to
n
al[pha]=n al=0.0038433
.00370 to
.0039999
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9999 sa: 1
Set serial sampling setting to n
seconds
sa[mple]=n sa=0 0 to 4000
Set serial duplex mode: du[plex]=f[ull]/h[alf] FULL or HALF
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f] ON or OFF
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of