Calibration Bath
Specifications and Environmental Conditions
3
COLD BATHS
Turning the Product on and off frequently adds wear to the compressor. Bath or
compressor lifetime can be affected by turning the Product off and on frequently.
Fluke Calibration recommends to turn the bath off if it is not going to be used for
the weekend or an extended period of time.
Specifications and Environmental Conditions
Safety specifications are listed in the Safety Information document that ships with the Product.
Note
The Product can show some control sensitivity to moderate or severe electromagnetic
fields or conducted interference of certain frequencies. In the presence of radiated EM
disturbances, with frequencies of 250 MHz to 400 MHz and with amplitude >1 V/m to a
maximum of 3 V/m, add 0.0025 °C to the stability specification. Stability is not guaranteed
if amplitude is >3 V/m. When subject to conducted disturbances of 8 MHz to 80 MHz, and
amplitude >3 V, add 0.005 °C to the stability specification.
Stability (2 sigma) ........................................... ±0.006 °C at –80 °C (ethanol)
±0.00C atC (ethanol)
±0.005 °C at 100 °C (5012 oil)
Uniformity .......................................................... ±0.007 °C at –80 °C (ethanol)
±0.00C atC (ethanol)
±0.007 °C at 100 °C (5012 oil)
Temperature Range ....................................... 5 °C to 35 °C (41 °F to 95 °F)
Mains Voltage ..................................................within ±10 % of nominal
Heating Time
[1]
................................................. 60 minutes, from 25 °C to 100 °C (5012 oil @ 230 V)
Cooling Time ..................................................... 210 minutes, from 25 °C to –80 °C (ethanol)
Stabilization Time ........................................... 15 minutes to 20 minutes
Temperature Setting...................................... Digital display with push-button entry
Set-point Resolution...................................... 0.01 °; 0.00018 ° in high resolution
Display Temperature Resolution............... 0.01 °
Digital Setting Accuracy .............................. ±1 °C
Digital Setting Repeatability ...................... ±0.01 °C
Wetted Parts ..................................................... 304 Stainless Steel
[1] Rated at listed 230 V.
Installation
The subsequent sections provide more detail on Bath installation.