EasyManua.ls Logo

Fluke ScopeMeter 190 Series III - Troubleshooting Common Issues; Resolving Test Tool Shutdown Issues; Troubleshooting Black Screen Problems; Handling Test Tool Software Hang-ups

Fluke ScopeMeter 190 Series III
97 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ScopeMeter
®
Test Tool
Maintenance
87
Voltage Probe Calibration
To fully meet user specifications, you need to adjust the voltage probes for optimal response. The
calibration consists of a high frequency adjustment and a dc calibration for 10:1 probes and
100:1 probes. The probe calibration matches the probe to the input channel.
To calibrate the 10:1 voltage probes:
1. Press
A to show the input A key labels.
2. Press 3 to open the PROBE ON A menu.
If the correct probe type is already selected (yellow shaded) you can continue at step 5.
3. Use the cursor and E to select Probe Type: Voltage and Attenuation: 10:1.
4. Press 3 to open the PROBE ON A menu again.
5. Press 1 to select PROBE CAL...
Note
It is necessary to connect both the hook clip and the zero reference contact.
A message appears asking you whether to start the 10:1 probe calibration.
6. Press 4 to start the probe calibration.
A message appears telling you how to connect the probe. Connect the red 10:1 voltage probe
to input A and to the probe calibration reference signal. See Figure 48.
Figure 48. Voltage Probe Adjustment
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com

Table of Contents

Related product manuals