FunctionPart
Column valve10
UV monitor11
Available modules
The modular design allows the user to customize ÄKTA pure in multiple ways. The system
is always delivered with the core modules of the selected configuration, but optional
modules may be added to the flow path.
The table below lists the available modules forÄKTA pure 25 and ÄKTA pure 150. Core
modules are indicated with an asterisk (*).
The valves for ÄKTA pure 25 and ÄKTA pure 150 are compatible with both
systems but for the best performance the specific valve type should be used.
The narrow channels in the valves for ÄKTA pure 25 will give too high back
pressure if used above 50 ml/min. The larger volumes in the “H” valves for
ÄKTA pure 150 may decrease resolution and increase peak broadening if used
in ÄKTA pure 25.
Note:
Label inModule
ÄKTA pure 150ÄKTA pure 25
P9H AP9 ASystem pump A*
P9H BP9 BSystem pump B*
R9R9Pressure monitor*
M9M9Mixer*
V9H-InjV9-InjInjection valve*
V9H-IAV9-IAInlet valve A
V9H-IBV9-IBInlet valve B
V9H-IABV9-IABInlet valve AB
V9H-IXV9-IXInlet valve IX
V9H-ISV9-ISSample inlet valve
V9H-MV9-MMixer valve
V9H-LV9-LLoop valve
42 ÄKTA pure Operating Instructions 29022997 AF
3 System description
3.1 ÄKTA pure instrument overview

Table of Contents