4.1.4 Power requirements
Introduction
This section describes the power supply requirements for ÄKTA pure.
Electrical power requirements
The table below specifies the electrical power requirements.
RequirementParameter
100 to 240 V AC ±10%Supply voltage
50/60 HzFrequency
Overvoltage category IITransient overvoltages
300 VA in state RunTypical power consump-
tion
165 VA in state Ready
25 VA in state Power-save
1100 VAMax power consump-
tion
1 socket per instrument, up to 3 sockets for computer
equipment
Number of sockets
EU or US plugs. Grounded mains sockets, fused or protect-
ed by equivalent circuit breaker.
Type of sockets
Maximum 2 m from the instrument (due to length of mains
cable). Extension cables can be used if required.
Location of sockets
WARNING
Protective ground. The product must always be connected to a
grounded power outlet.
64 ÄKTA pure Operating Instructions 29022997 AF
4 Installation
4.1 Site preparation
4.1.4 Power requirements