Laboratory bench
The bench must be clean, flat and stable and able to support the weight of ÄKTA pure,
see table below.
Equipment dimensions
630 mm
470 mm
535 mm
Equipment weight
WeightItem
up to 53 kgÄKTA pure instrument
approximately 9 kgComputer
approximately 3 kgMonitor
ÄKTA pure Operating Instructions 29022997 AF 61
4 Installation
4.1 Site preparation
4.1.2 Room requirements