Prime the pump rinsing system
Follow the instructions below to fill the pump piston rinsing system with rinsing solution.
See the tubing configuration of the rinsing system in the illustration above.
ActionStep
Remove the pump rinsing liquid tube from the holder.1
Fill the pump rinsing liquid tube with 50 ml of 20% ethanol.2
Place the pump rinsing liquid tube back in the holder.3
Insert the inlet tubing to the system pump piston rinsing system into the
fluid in the rinsing solution tube.
Note:
Make sure that the inlet tubing reaches close to the bottom of the rinsing so-
lution tube.
4
Connect a 25 to 30 ml syringe to the outlet tubing of the system pump piston
rinsing system. Draw liquid slowly into the syringe.
5
84 ÄKTA pure Operating Instructions 29022997 AF