Step Action
3 Hang the air trap tubing on the valves on the front of the system as illustra-
t
ed below. Then position the tubing so that it is located on the opening side
of each valve.
Note:
I
f a filter is needed in the planned process, the filter should be fitted after the
installation test because the installation test is performed on a dry flow kit.
See Section 5.5 Install filter (optional), on page 129 for instructions.
4 Locate the tubing size adapter (reducer).
Note:
T
he tubing size adapter is only used for the isocratic setup.
5 Preparation
5
.2 Install disposable flow kits
108
ÄKTA ready XL Operating Instructions 29281616 AE