Step Action
5 For the isocratic setup, connect the air trap tubing (A) to the pump cham-
b
er outlet labeled OUT with a TC clamp with the tubing size adapter (B).
The illustration above shows the isocratic setup.
6 For the gradient setup, connect the air trap tubing to the jumper tubing
w
ithout using the tubing size adapter.
5 Preparation
5
.2 Install disposable flow kits
ÄKTA ready XL Operating Instructions 29281616 AE 109