Step Action
7 Connect the C
E-101/TE-101 sensor to the the cable connector located on
the air trap flow kit leading to the pump chamber.
Secure the connection by twisting the locking ring clockwise on the cable
connector.
The illustration shows the isocratic setup.
8 Unlock the front wheels of the pump wagon.
9 Push in the pump wagon all the way in under the system and lock it in place
w