Emergency shutdown
A
A
B
ResultActionStep
Press one of the EMERGENCY
STOP buttons (A).
1.
All motors and externally moving
components stop immediately.
The built-in computer and other
components remain powered.
No data is lost.
The valves will shut down in a specific
order, for details see Valve shutdown
order below.
If required, also switch off the
mains power supply using the
SYSTEM POWER SWITCH (B),
disconnect the power cord or
switch off the fixed power
supply circuit breaker.
2.
Power to the entire system, including
the computer, is lost.
Data and run status may be lost.
All valves will shut immediately and not
in sequence.
ÄKTAprocess Operating Instructions 29-0252-49 AA 35