5.2 Perform a run
Precautions
WARNING
Cabinet doors. During operation, all doors must always be closed
and locked.
WARNING
Use a harmless fluid in the beginning of the process. This will make
it possible to detect leakage with minimized consequences and
the risk for potential leakage of hazardous fluids is avoided.
CAUTION
Use ear protection whenever working close to the system in
operation.
NOTICE
Excessive temperatures may damage the equipment. Do not run
the system at higher temperatures than the specified maximum
operation temperature as stated on the system label.
NOTICE
Operator console
Do not use the operator console to push or drag the system
Do not lean on the console
The console arm is only designed to support the weight of the
operator console.
124 ÄKTAprocess Operating Instructions 29-0252-49 AA