Programming Site Data 2-5
September 2004
Entering Transducer and
Pipe Parameters - CHx-
PIPE
Enter the transducer and pipe parameters via the PIPE submenu. While
following the programming instructions, refer to Figure A-1 on page
A-1 in Appendix A, Menu Maps. Remember to record all
programmed data in Appendix C, Data Records.
Accessing the PIPE
Submenu
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [
X] until PROG appears and press [ENTER].
3. Press [
X] until the desired channel appears. Press [ENTER] at CH1
or CH2.
4. Press [
X] until PIPE appears and press [ENTER].
5. Do one of the following:
• Standard transducers, enter the transducer number by using the
arrow keys. Press [
W] and [X] to move the cursor to desired
location and press
[S] and [T] to increase or decrease the
number and press [ENTER]. Proceed to the next step.
• Special transducers, proceed to Special Transducers below.
IMPORTANT: Special transducers have no engraved number on the
head and are rarely used. Examine the transducer
head carefully for a number.
6. Go to one of the following sections:
• Standard clamp-on transducers - proceed to Pipe Material on
page 2-7.
• Standard wetted transducers - proceed to Pipe Outside
Diameter on page 2-8.