Programming Site Data 2-17
September 2004
Initializing Setup
Parameters - INIT
Use this option to initialize (reset) all of the parameters within the SET
UP
menu back to their default values. Complete the following steps to
reset all of the parameters:
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [
X] until PROG appears and press [ENTER].
3. Press [
X] until the desired channel appears. Press [ENTER] at CH1
or CH2.
4. Press [
X] until SETUP appears and press [ENTER].
5. Press [
X] until INIT appears and press [ENTER].
6. Select YES or NO.
Procedure Options After completing the above steps, the user program returns to the
SET UP window. Do one of the following:
To continue regular programming, refer to Appendix A, Menu
Maps, to navigate to the desired menu.
To leave the user program, press [ESC] four times.
Using Advanced Features
- ADVAN
This option enables you to access the more advanced features of the
meter. In this option you can do the following:
KV/SS - calculate the kinematic viscosity (KV) based on signal
strength (SS) - refer to page 2-18.
MULTK - enter a table of K-factors (based on velocity or
reynolds number) that compensates for non-linear flow rates -
refer to page 2-18.
MASS - enable mass flow (calculated for static fluid density) -
refer to page 2-20.
CODEL - select the size of the transducer transmission signal -
refer to page 2-21.

Table of Contents

Other manuals for GE AquaTrans AT868

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AquaTrans AT868 and is the answer not in the manual?

GE AquaTrans AT868 Specifications

General IconGeneral
BrandGE
ModelAquaTrans AT868
CategoryTransmitter
LanguageEnglish

Related product manuals