September 2004
2-34 Programming Site Data
Configuring the Output
as a Totalizer or
Frequency Output
Output B and/or D can be configured as a totalizer or frequency
output using Instrument Data Manager software. Before the output
can be configured, the AT868 must be connected to a PC using the
serial port (see Wiring the Serial Port on page 1-9).
Use the following steps to configure the output:
1. Open the SITE EDIT MENU from the CALIBRATION/TEST menu.
2. Select Hardware Test.
3. Select SET TOT/FREQ.
4. Use the pull-down menu to select the desired output (frequency or
totalizer) and press Next Item/Enter.
5. To leave the user program, press Exit Page twice.