September 2004
2-8 Programming Site Data
Pipe Outside Diameter The programming sequence should be rejoined here for all
transducers.
1. Press [
X] to move the cursor to the outside diameter units and
press
[S] and [T] to select the desired units and press [ENTER].
2. Enter the known pipe outside diameter or circumferences by using
the arrow keys. Press [
W] and [X] to move the cursor to desired
location and press
[S] and [T] to increase or decrease the
number and press [ENTER].
Obtain the required information by measuring either the pipe
outside diameter (OD) or circumference at the transducer
installation site. The data may also be obtained from standard pipe
size tables found in Sound Speeds and Pipe Size Data manual
(914-004).
3. Enter the know thickness of the pipe wall by using the arrow keys.
Press [
W] and [X] to move the cursor to desired location and
press
[S] and [T] to increase or decrease the number and press
[ENTER]. If the pipe wall thickness is not available, look up the
value in a table of standard pipe size data which can be found in
Sound Speeds and Pipe Size Data manual (914-004).
4. Do one of the following:
• All wetted transducers - proceed to Path and Axial Length
below.
• All clamp-on transducers - proceed to Lining on page 2-9.
Path and Axial Length 1. Enter the path length of the ultrasonic signal.
2. Select the desired units and press [ENTER].
Note: If a spoolpiece was ordered with the meter, the transducer
signal path length
(P) and the transducer signal axial length (L)
are engraved on the flowcell and/or are included in the
documentation supplied with the meter. For on-site transducer
installations, refer to Appendix D, Measuring P and L
Dimensions, for instructions.
3. Enter the axial length of the ultrasonic signal and press [ENTER].
4. Select the desired units and press [ENTER].
5. Proceed to Setting Up Tracking Window on page 2-9.