GE AquaTrans AT868 - Resetting Totals

GE AquaTrans AT868
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Displaying Data 3-3
September 2004
Resetting Totals To reset the totals, you can use the internal keypad, PanaView™ or an
external switch. When resetting totals, the totals for both channels
will be reset. Use the appropriate section that follows to reset
totalizers.
Resetting Totalizers Using
the Internal Keypad
Refer to Figure A-4 on page A-4 in Appendix A, Menu Maps.
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [
X] until RESET appears and press [ENTER].
3. Press [
X] to select YES or NO and press [ENTER].
If you select YES, the AT868 resets the totals for both channels. If you
select NO, the totals continue.
4. To leave the User Program, press [ESC].
Resetting Totalizers Using
PanaView™
Refer to Figure F-4 on page F-4 in Appendix F, PanaView™ for
IDM-Combatible Meters.
1. Open the System menu.
2. Select Clear Instrument Totalizers.
3. PanaView™ automatically clears the totalizers and exits the
System menu.
Resetting Totalizers Using
the External Switch
While the meter is operating, press and hold the reset switch for 1
second. Totals will reset to zero and immediately begin to accumulate
again.
Note: The external totalizer reset switch must be configured as
described on page 2-23.