Programming Site Data 2-11
September 2004
KV Input Selection 1. Press [W] and [X] to enter a static kinematic viscosity or a table
of values and press [ENTER].
2. Do one of the following:
• If you selected Table, enter the Calibration Factor by using the
arrow keys. Press [
W] and [X] to move the cursor to desired
location and press
[S] and [T] to increase or decrease the
number and press [ENTER].
Note: The values in the KV/SS table can be edited as described in
Entering KV/SS Table on page 2-18.
• If you selected STATC, the AT868 will select and automatically
display the Kinematic Viscosity (see note below). If you want
to enter a different number use the [
W] and [X] to move the
cursor to desired location and press
[S] and [T] to increase
or decrease the number and press [ENTER]. Then enter the
Calibration Factor in a similar manner and press [ENTER].
Note: If you select OTHER as the fluid type, you must enter the
Kinematic Viscosity. The AT86 8 will only approximate the
Kinematic Viscosity if you select one of the available fluid
options other than OTHER (i.e. water, oil, etc.).
3. Proceed to one of the following sections:
• For Clamp-on Transducers - proceed to Number of Traverses
and Transducer Spacing on page 2-12.
• For Wetted Transducers - The meter returns to the Channel
PROGRAM window. At this point, you can continue
programming in this submenu as desired. To leave the Channel
PROGRAM window, press [ESC] three times. To complete
setting up the meter, you must enter data in the GLOBL-SYSTM
submenu as described on page 2-22.