8 Troubleshooting
86500-IMG rev 3 GE Healthcare 63
Er0075 Internal communication
error. Compression
Board not answering.
Clear and try again. If
problem persists, call service.
Er0076 Internal communication
error. Stereo Board not
answering.
Clear and try again. If
problem persists, call service.
Er0077 Internal communication
error. Collimator Board
not answering.
Clear and try again. If
problem persists, call service.
Er0078 Internal communication
error. Control Panel not
answering.
Clear and try again. If
problem persists, call service.
Er0080 Stereo movement back
up time reached.
Clear and try again. If
problem persists, call service.
Er0081 Cassette in rear position. Clear and continue use.
Should cassette not move to
the front position, drive it
there manually.
Er0082 1) C-Arm rotation with
the tube head in rear
position disabled when
vertical down C-arm
angle over 120°.
2) Z-movement down
blocked when the C-arm
angle exceeds 122°.
1) For C-Arm angles over
120°, drive the tube head to
the front position. For lowest
Z-position, drive drive the
tube head is in front position.
2) Drive the C-arm to less than
122° (the tube head must be
in front position).
Er0084 Transmit buffer overflow
in Supervisor software.
Clear and try again. If
problem persists, call service.
Release compression
manually.
Er0087 Movement inhibited
during compression.
Release compression and try
again.
Er0089 C-Arm center limit not
found.
Drive manually in the center.
If problem persists, call
service.
Er0090 Tube not in front
position.
Remove Delta 32 or biopsy
paddles to enable tube head
AutoDrive.
Er0091 Z-movement disabled in
Specimen mode.
Clear and continue. If a
patient is to be examined, exit
the Specimen mode.
Error code
in LCD
display
Reason for error Corrective action
Table 8.1 Diamond error code guide