8  Troubleshooting
62 GE Healthcare 86500-IMG   rev 3
Er0059 Anode run failure. Clear and try again. If 
problem persists, call service.
Er0060 Exposure button 
released during 
exposure.
Load new film and expose. 
Hold exposure button down 
until the audio signal ends.
Er0062 Generator failure. kV not 
OK.
Clear and try again. If 
problem persists, call service.
Er0063 Line voltage low. Have an electrician check the 
voltage of the electric outlet. 
If it is strong and stable 
enough, then call service for 
further actions.
Er0064 Cassette missing. Insert a cassette and try also 
with another type.
Er0065 Exposure time exceeds 
limit (too long/too short).
Increase/decrease kV values 
and try again. If problem 
persists, call service.
Er0066 Movement button or 
switch exposure button 
pressed during 
movement or start-up, 
function not allowed.
Clear and try again. If 
problem persists, call service.
Er0067 Device not ready for 
exposure.
Clear and try again. If 
problem persists, call service.
Er0068 Delta 32 PC refused to 
receive exposure.
Clear and try again. If 
problem persists, call service.
Er0069 Delta 32 PC connection 
time-out.
Clear and try again. If 
problem persists, call service.
Er0070 Error in stereo motor 
pulse measuring.
Clear and try again. If 
problem persists, call service.
Er0071 Tube head heat limit 
exceeded.
Wait and let cool down 
enough. Remaining cooling 
time indicated in timer screen. 
Clear and try again. If 
problem persists, call service.
Er0072 Side collimator failure. Clear and try again. If 
problem persists, call service.
Er0074 Internal communication 
error. AEC Board not 
answering. Cable 
broken.
Clear and try again. If 
problem persists, call service.
Error code 
in LCD 
display
Reason for error Corrective action
Table 8.1 Diamond error code guide
FOR TRAINING PURPOSES ONLY! 
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).