10 Specifications
86500-IMG rev 3 GE Healthcare 73
SF = Small focus, LF = large focus
Time range with loading factors shown above:
Focus 0.3 0.02 - 5.0 sec.
Focus 0.1 0.04 - 10.0 sec.
mAs range Focus 0.3 mm 2 - 500 mAs:
Tolerance:
mAs 40: ± 4 mAs
mAs > 40: ± 10%
Focus 0.1 mm 4 - 300 mAs
Tolerance:
mAs 40: ± 4 mAs
mAs > 40: ±10%
Time range Focus 0.3 mm 0.02 - 5 s
Focus 0.1 mm 0.04 - 10 s
Tolerance: ± 0.10 + instrument
inaccuracy
Tube loading chart for Diamond X-ray tube (M113SP)
kV M113SP
mA(SF)
M113SP
mA(LF)
kV M113SP
mA(SF)
M113SP
mA(LF)
15 18 53 28 25 85
16 20 59 29 24 85
17 22 64 30 23 80
18 23 69 31 22 80
19 24 74 32 21 80
20 25 78 33 21 75
21 26 81 34 20 75
22 28 85 35 20 69
23 30 90 36 19 50
24 29 95 37 19 50
25 28 100 38 18 50
26 27 95 39 18 50
27 26 90
Generator
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).