10 Specifications
74 GE Healthcare 86500-IMG rev 3
1 Focal spot is marked in the front part of the tube head assembly.
2 The collimator label is at the front end of the tube head assembly.
Automatic Exposure Control (AEC)
Solid state 15 step density selector
Operating range for large focus 23 - 35 kV
for small focus 25 - 35 kV
Nominal shortest irradiation time 60 msec.
X-RAY TUBE Diamond (Varian M113SP in B112
housing)
Anode type Rotating dual-angle anode
Anode heat capacity 300,000 HU
Target material Molybdenum, niobium doped
Anode rotation speed 3000rpm(50Hz)/3600rpm(60Hz)
Focal spot sizes M113SP 0.1, 0.3 (IEC60336)
Port material beryllium
Reference axis 42 mm (M113SP) from chest wall
edge (tube angle 4°)
Diamond Filtration
Molybdenum
Rhodium
Aluminum
0.025 mm
0.025 mm
0.50 mm
Positioning
Source to Image Distance (SID) 660 mm (26”)
Movement of the motorized, dual-speed carriage
Rotation
Vertical
±185°
760 mm (30”)
The cassette holder vertical movement range is 68 cm - 144 cm above
finished floor level (27” - 57”) with the C-Arm in the upright position (0°)
and 79 cm - 156 cm (31” - 61”) with C-Arm rotated in the lateral (90°)
position.
(26”)
(66cm)
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).