10 Specifications
86500-IMG rev 3 GE Healthcare 75
Compression
motorized
manual
Max 200N
Max 270N±20N
Foot pedal protection
classification
IPX1
Recording systems
Bucky for 18 x 24 and 24 x 30cm film
Magnification for 18 x 24 cm film 1.6, 1.8 & 2.0
exposed size 10 x 24
Stereotactic device Delta 32 digital ccd-camera
Display accuracy
Compression force ± 20N
Compression thickness ± 1cm
C-Arm angle ± 2°
mGy display ± 30%
For mGy display accuracy, refer to:
“Spectral Dependence of Glandular Tissue dose in Screen-Film
Mammography” Wu Xizeng, Wu X, Barnes Gary T, Tucker Douglas M,
Radiology 1991; 179:143-148
“Normalized Average Glandular dose in Molybdenum Target- Rhodium
Filter and Rhodium Target -Rhodium filter Mammography” Wu Xizeng,
Wu X, Gingold Eric L, Barnes Gary T, Tucker Douglas M, Radiology 1994;
193:83-89
Correspondence between X-ray field and image reception area
Chest wall 0...+3 mm
Other walls +1...+5 mm
Correspondence between
collimator light field and X-ray
field
5 mm any wall
FOR TRAINING PURPOSES ONLY!