Chapter 1. Programming Site Data
28 DigitalFlow™ GF868 Programming Manual (1-Channel)
1.10.4 Setting Advanced Parameters (cont.)
1.10.4b Edit K Factors
1. Press [F2] = MULTK to enter the MultiK option.
2. Press [F1] = NO to disable or [F2] = YES to enable Multiple K-factors.
If you press NO, the GF868 returns to the Advanced Features prompt.
If you press YES, complete the steps below.
a. The GF868 asks for the type of K factor. Press
[F1] to select CstV (Custom velocity) or [F2] to select CstR
(Custom Reynolds number).
If you enable CstR, you must also enter the Kinematic Viscosity of your gas, as listed in Sound Speeds and Pipe
Size Data. Use the numeric keys to enter a value, and press [ENT].
b. Press
[F1] = NO to retain the current K-factor table and return to the Advanced Features prompt, or press [F2] =
YES to Edit The K-factor Table.
Note: If velocity vs. K-factor data was not provided with the Model GF868, the K-factor table can not be edited.
c. Enter the Number of K-factors to be entered into the table. Press [ENT]. (The meter will only accept values from
2 to 20.)
IMPORTANT: When editing the K-factor table, the velocities must be entered in increasing order.
d. Enter the Velocity (or Reynolds) Value for K-factor number “x”. Press [ENT]. (The meter will only accept values
from –30,000 to +30,000 ft/sec.)
e. Enter the K-factor corresponding to velocity number “x”. Press [ENT].
3. Repeat steps c and d until all of the data points have been entered.
1.10.4c Procedure Options
After completing the above steps, the meter returns to the Advanced Features prompt. Continue as follows:
To continue programming the meter, refer to the menu maps in Appendix A and navigate to the desired menu.
Then, proceed to the appropriate section of this manual for instructions.
To leave the User Program and retain the previous settings, press [EXIT] twice and then press [F1] = NO at the SAVE
prompt. Any programming changes will be discarded and you will be returned to the data display.
To leave the User Program and return to measurement mode, press [EXIT] twice and then press [F2] = YES at the
SAVE prompt. Your programming changes will be entered into the meters memory, and you will be returned to
the data display.