DigitalFlow™ GF868 Programming Manual (1-Channel) 55
Chapter 3. Logging Data
3.4 Stopping a Log
Use the STOP submenu to terminate a logging process that is currently active. Press the appropriate side of the [SCREEN]
key to activate the desired pane of the display screen, and complete the following steps:
1. To access the Log Menu, press the [LOG] key.
a. If you have activated the security feature (see Chapter 1, Programming Site Data, page 35 for details), enter the
assigned password and press [ENT].
Note: Once a log is stopped it cannot be restarted, but the log remains in memory. To clear the log from memory,
refer to Chapter 5, Clearing Data, for instructions.
2. Press [F3] = STOP.
3. Use the [, [ and [F1]-[F4] keys to select a log file to stop.
Note: At the NAME prompt, the option bar will show the names of all completed or active log files that are currently
stored in memory. Recall that a flashing asterisk (*) appears at the far right of the locator bar if the Model
GF868 is currently logging data (see Chapter 3, Operation, of the Startup Guide).
4. Press [F1] to continue logging and return to the initial Log Menu screen. Press [F2] to stop logging and return to the
initial Log Menu screen.
3.4.1 Procedure Options
Do one of the following:
Press [F1]-[F4] to enter another LOG submenu.
Press [EXIT] to return to the data screen.