Electrical Requirements 5.5 System Cable Information
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 5-13
Injector
Console
(adaptor box)
Y 8.2 2.5 Power cable injector to console
ECG CT gantry N 30 9.1 Gantry to ECG monitor
NM gantry CT gantry N 126 29.5 9 CT table interface cable
NM gantry CT gantry N 124 29.5 9 NM-CT gantry interface cable
NM gantry CT gantry N 125 29.5 9 ECG and respiratory gating cable
NM gantry Patient table N 130 65.6 (62.3) 20 (19)
Table cable harness
‡
NM gantry
PIDB
†
Y 120 65.6 (54.1) 20 (16.5) NM mains cable
MDP PIDB Y 9 2.5 PIDB-to-MDP power feed
NM gantry NM host N 128 65.6 (54.1) 20 (16.5) NM gantry to Host Ethernet cable
NM gantry NM host N 127 65.6 (54.1) 20 (16.5) NM gantry to Host persistence cable
* Actual length is the entire cable length, connector to connector, including cable run inside each sub-system.
Usable Length is the portion of the cable routed externally between sub-systems.
† The MDP can be the A1 or other primary power disconnect device
‡ The standard table cable provided is 20 M. There is an optional 10 M cable available.
Table 5-5: D670 Sub-system Inter-connection Cables(Continued)
Start / Destination
H/V
Separation
(Y or N)
Run
#
Actual Length
(Usable Length)
*
Description
From To
Feet Meters