Chapter 6. REFERENCE AND SPECIFICATION
100 | ADTS 405 Mk2 Instruction Manual-English © 2015 General Electric Company – All rights reserved.
ADTS 405F and ADTS 405R Performance expressed in example pressure units
Control Performance
Stability
The ADTS 405 provides stable pressure values at the outlet ports and continues to control
smoothly without oscillation into volumes of up to 17 litres (1,000 cubic inches) on Ps and a volume
of up to 10 litres (600 cubic inches) on Pt at rates up to 6000 ft/min. Lower volumes enable higher
rates of change to be achieved.
The following are the figures based on a 40 ppm of Ps full-scale or Qc full-scale over a bandwidth of
0.1 to 1.5 Hz and using RMS values. Below are shown the two Qc ranges used by the ADTS 405
Static
(Ps)
Rate Ps Pitot
(Pt)
1000
knts
Pitot
(Pt)
850
knts
Differential
(Qc)
1000
knts
Differential
(Qc)
850
knts
Rate Qc EPR
Units mbar
abs
mbar/
min
mbar abs mbar abs mbar diff mbar diff mbar/
min
-
Maximum
Range
a
a. Pressures below 35 mbar absolute and the peak rates of change are available but subject
to the use of a suitable vacuum pump.
3 to
1355
10,000 3 to 3500 3 to 2700 -1352 to
3377
-1352 to
2700
10,000 0.1 to 10
Calibrated or
Standard
Range
a
35 to
1355
200 35 to
3500
35 to
2700
-1320 to
2500
-1320 to
1700
200 0.1 to 10
Accuracy
b
b. Accuracy, (expanded uncertainty K=2); includes Non-linearity, Hysterisis and repeatability
over full temperature range, 12 months drift and calibration standard uncertainty.
±0.1 ±1%
within 5
seconds
0.26mbar 0.20mbar RSS of Ps
and Pt
c
c. The accuracy figures assume Auto-Zero function is off.
RSS of Ps
and Pt
c
±1% better
than 0.005
Resolution 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.001
Channel Controller
Stability
Ps 1,355 mbar
40 inHg
±0.038 mbar
±0.0011 inHg
Qc 1,700 mbar
50 inHg
±0.045 mbar
±0.0014 inHg
Qc 2,500
74 inHg
±0.070 mbar
±0.0021 inHg

Table of Contents