ADTS 405R/F Specification
© 2015 General Electric Company – All rights reserved. ADTS 405 Mk2 Instruction Manual-English | 97
6.3 ADTS 405R/F Specification
Rack mounted ADTS 405R (indoor use):
Size Standard 19" rack front panel 6U high (10½”)
Depth behind front panel 255 mm (10")
Rear panel connections available
Weight 12 kg [26.5 lb]
EMC EN 61326-1 (refer to the declaration of conformity for details)
Electrical safety EN61010-1
Power supply Single phase AC in the range
or
100/120/230V ~ 50/60 Hz
115V ~ 400 Hz
400VA
No switching required All automatic
Transient over-voltages up to
the levels of Overvoltage Cat-
egory
II
Operating altitude Up to 15,000ft (4,570 metres)
Operating temperature -20°C to 50°C (-4° to 122°F)
Storage and shipping temperature -51°C to 70°C (-60° to 158°F)
Ingress Protection IP23
Operating humidity 0 to 95% RH (non-condensing)
Pollution degree 3
Pressure safety:
Pressure Equipment Directive - class: Sound engineering practice (SEP)
Shock and vibration
ETSI EN 300 019-2-5 Class 5.1 Ground Vehicle
Installations.
Test method: EN 60068-2-64 Random Vibration
Pressure Fittings Ps MS33656-6
Pt MS33656-4
Pressure supply MS33656-4
Vacuum supply MS33656-6