Introduction
© 2015 General Electric Company – All rights reserved. ADTS 405 Mk2 Instruction Manual-English | 77
6.2.14 [DISPLAY QNH]
1. When [DISPLAY QNH] is selected, the local atmospheric pressure recorded during power-
up is converted to the equivalent sea level pressure. The display shows QNH, in the default
units set in configuration, together with the station (airfield) altitude.
2. The station (airfield) altitude can be changed in full set-up mode by using SETUP,
GROUND.
6.2.15 PORT
This function sets the port option, this changes the ADTS 405 from a single pitot-static output
system to a multi-port system by using a Druck Line Switching Unit (LSU 100 Series). Refer to the
option manual.
6.2.16 REMOTE
Enters Remote Mode; this mode allows an external IEEE 488 bus device to control the unit. In this
mode the key-pad (except the Remote key) is disabled. If remote lock is disabled, pressing the
remote key when the “REMOTE” message is not displayed forces the unit into remote mode as if
the external device has taken control of the unit.
Note: This may change operator settings such as the current limit set. Refer to
the option user manual for further details.
6.2.17 PRINT
The ADTS 405 contains a print facility to print the displayed readings together with up to nine user
readings (e.g., Pilot, Copilot, Aux) to a text file stored within the ADTS internal file storage system
which may be accessed and printed via PC.
Access to the stored 'Print' files area within the ADTS is gained via the Ethernet port, using a PC to
create a secure file transfer connection. Details of how to set up an Ethernet connection are given in
section 3.7.
1. This key causes a printout of the date and time followed by the measured value of the
parameters. A prompt asks for the entry of a reading which can consist of any characters A
to Z, 0 to 9. (decimal point) or - (minus), in any order. It can be used to enter the value from
the unit or system under test or alternatively a test number. When no more entries are
required, press CLEAR/QUIT.
2. To enter an alphabetic character (A to Z), press the [ALPHA] key and use theor
(nudge keys) to change to the required character. To enter another alphabetic character,
press the [ALPHA] key again. Press [BACK] to delete the last character.
3. Up to nine user readings can be entered sequentially. The display prompts for the next user
reading each time one is entered. When no further readings are required, press CLEAR/
QUIT. Example print output:
04Mar00 09:47 Ps 700.00 mbar Qc 30.00 mbar Usr Rdg

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Druck ADTS 405 R/F Mk2 and is the answer not in the manual?

GE Druck ADTS 405 R/F Mk2 Specifications

General IconGeneral
BrandGE
ModelDruck ADTS 405 R/F Mk2
CategoryTest Equipment
LanguageEnglish

Related product manuals