4. MAINTENANCE
4.1 Introduction
This section details the recommended maintenance tasks to be carried out by the operator. The
maintenance chart shows the periodicity of each task and a code referenced to the task detailed in
4.3 Maintenance Tasks.
Table 4-1: Maintenance Chart
* Periodicity may change depending on usage and environment, refer to the
Engineering Authority.
4.2 Materials
This lists materials required for the user to maintain the ADTS 405. The item number and name in
the table identifies the item in the maintenance tasks (for example, “Use a soft bristle brush, item 4,
Table 4-2”). The description provides further information required to identify the item.
Table 4-2: Materials List
Note: Equivalent substitutes can be used.
Task Code Period
Inspect A Daily, before use
Inspect B Weekly
Test C Before use
Test D Daily, before use
Calibrate E Every 12 months *
Replace F
As detailed in fault finding or when detailed after
inspection
Clean G Weekly *
Service H
1,000 hours recorded by the pump unit elapsed time
indicator or the equipment running log
Service J
3,000 hours recorded by the pump unit elapsed time
indicator or the equipment running log
Item
Number
Item Name, Description
1 Cloth, Cotton, Lint-free
2
Alcohol, Isopropyl (MIL-A-10428,
Grade A)
3 Detergent, Mild, liquid
4 Brush, soft bristle (MIL-B-43871)