Chapter 3. OPERATION
28 | ADTS 405 Mk2 Instruction Manual-English © 2015 General Electric Company – All rights reserved.
3.5.1.2 Change the display for Pt Only:
a. Press SETUP.
b. Select [MORE].
c. Select [Control], [Ps Pt Dual].
d. Select [Pt Only Mode].
e. Press CLEAR/QUIT repeatedly to return to user display.
3.5.1.3 Change measurement units:
In Full Set-up Mode any set of aeronautical or pressure units can be selected.
a. Press the SETUP key.
b. Select [UNITS].
c. Use [NEXT] and [PREV] keys to select the required units.
d. Select [SAVE] to accept.
e. Press CLEAR/QUIT repeatedly to return to the main pressure display.
In Minimum Setup Mode only the default aeronautical and pressure units can be
selected.
If Set-up Mode is switched off, units cannot be changed.
3.5.1.4 Change the unit type
a. Press the SETUP key.
b. Select [UNITS TYPE].
c. Select either [AERO] or [PRESS] to select the unit type.
d. Select [SAVE] to accept.
e. Press CLEAR/QUIT repeatedly to return to the main pressure display.
The default pressure and aeronautical units are defined in the CONFIGuration
mode.
Display cannot be changed with SETUP switched off or in minimum mode.
3.5.2 Aim
The ADTS 405 must be in control mode to set a new aim (current aim displayed). If the aim field
shows “Leak Measure” press LEAK MEASURE/CONTROL to enter control mode.
A new aim can be entered using the numeric keys. The existing aim is replaced when the first digit
of the new aim is pressed. Each digit is displayed as it is pressed, If an error is made during the entry
of data, press CLEAR/QUIT to restore the original aim.
Press ENTER to action the new aim.
Note: The 000 key can be used as a quick way of entering thousands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Druck ADTS 405 R/F Mk2 and is the answer not in the manual?

GE Druck ADTS 405 R/F Mk2 Specifications

General IconGeneral
BrandGE
ModelDruck ADTS 405 R/F Mk2
CategoryTest Equipment
LanguageEnglish

Related product manuals