GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336113-1EN, REVISION 10 INSTALLATION MANUAL
Page 84 Section 4.10 Installing hardware options
10. Mount the clamp to the Thorax chassis. Slide the clamp to the rear of the mounting screw slots.
11. Tighten the clamp around the cable bundle, ensuring that the cables are not pinched by the
clamp.
Figure 4-14 Correct cable bundle arrangement
12. Install the front bin, top cover and side covers.
13. Re-install the cable cover on the vertical column cable bundle. See the System manual,
Replacement Procedures chapter, Cable Cover Installation section.
14. Configure the system to enable the DAP meter option. See 4.2.7 Options configuration on
page 64.
15. Once the DAP meter option has been enabled, it must be ’switched on’ at the Applications
level:
a. On the Applications desktop, click the UTILITIES
button.
b. On the Utilities desktop, select PREFERENCES
> OPTIONS.
c. Turn the DAP function ON
.
d. Refer to the system Operator Manual for instructions on DAP usage.
Item Description
1 DAP cable (optional for Brivo XR285amxOptima XR200amx systems)
2 Wireless cable (only valid for Optima XR220amx systems)
3 Vertical brake cable
4 Horizontal arm lock cable
1
2
C
3
4