3 Installation
This chapter provides information regarding installation of the ImageQuant LAS 4000.
3.1 Site requirements
The following table gives a summary of power supply and environmental requirements. For detailed
information, refer to Section 7.1 Main specifications, on page 79.
RequirementParameter
13 A, 100 - 120 V or 10 A, 200 - 240 VPower supply
50 to 60 HzLine frequency
15°C to 28°C, fluctuation < 10°C/hAmbient temperature
(Storage: -25°C to 70°C)
Stable laboratory bench, load capacity > 100kg/m
2
Placement
At least 20 cm free on all sides
30% to 70%, non-condensingHumidity
(Storage: 5% to 100%, non-condensing)
3.2 Transport
The ImageQuant LAS 4000 weighs approximately 57 kg and requires at least two people to lift and
move it.
Before moving the instrument:
1
Turn off the instrument.
2
Disconnect the power cord.
3
Remove the lens, transmitted light source and any loose parts from inside of the IDX. Refer to the
ImageQuant LAS 4000 User Manual.
3.3 Unpacking
When unpacking:
Check the equipment for any apparent damage before installing.
26 Getting Started with ImageQuant LAS 4000 28-9607-59 AB
3 Installation

Table of Contents

Other manuals for GE ImageQuant LAS 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ImageQuant LAS 4000 and is the answer not in the manual?

GE ImageQuant LAS 4000 Specifications

General IconGeneral
BrandGE
ModelImageQuant LAS 4000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals