6
Turn the filter changer, put a filter in the desired position and secure it in place.
7
Select the appropriate filter from the drop-down list and click the OK button.
Note:
The filter changer returns to the original position when you click the Start, Focusing or
Method/Tray position buttons.
3.5.5 Installing a transmitted light source
This section describes how to insert the UV transilluminator or the white light table into the IDX.
1
Push the transmitted light source along the guide at the bottom of the IDX.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB 39
3 Installation
3.5 Installation of hardware
3.5.4 Attaching a filter to the filter changer