CountermeasureCausePhenomenon
See the countermeasures for the SensKey and
SensCode numbers displayed on screen of the
analysis PC in Section 6.5 Message “ImageQuant
LAS 4000 hardware error occurred" , on page 69
A hardware error oc-
curred
ImageQuant LAS 4000 beeps on and
off (three times) and the ERROR LED
lights
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
ImageQuant LAS 4000
failure
ImageQuant LAS 4000 generates a
strange odor
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
ImageQuant LAS 4000
failure
ImageQuant LAS 4000 emits smoke
6.3 Problems with ImageQuant LAS 4000
If the error persists even after the following countermeasures have been taken, fill in the necessary
items and the error code in the Service report fax sheet at the end of the User Manual.
CountermeasureCausePhenomenon
Connect the power cable correctlyPower cable installation
failure
LED lights on ImageQuant LAS
4000 do not light during startup
diagnosis
Remove the foreign objectForeign object is present
in the locking section or
door sensor
The ImageQuant LAS 4000 door
cannot be opened and closed.
The door cannot be locked
Contact your GE Healthcare represen-
tative
The locking section is
damaged.
Turn the power off and check the
camera cable connection
The camera cable is not
connected correctly
The initial diagnosis does not
start even if the power is turned
on
Connect the AC power cable and
camera cable correctly
Power cable and cam-
era cable connection
failure
A focused image is not displayed
Connect the trans light source by
pushing it all the way until it hits the
other end. Lock the Epi lights in place.
Connector connection
failure. Light failure.
Remove the foreign objectForeign object is present
in the door sensor
Click the Focusing button in the soft-
ware
Software is not in the
Focusing mode
Getting Started with ImageQuant LAS 4000 28-9607-59 AB 67
6 Troubleshooting
6.2 Strange sounds or odors are perceived

Table of Contents

Other manuals for GE ImageQuant LAS 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ImageQuant LAS 4000 and is the answer not in the manual?

GE ImageQuant LAS 4000 Specifications

General IconGeneral
BrandGE
ModelImageQuant LAS 4000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals