Turn the filter changer so the desired position is easily accessible. If a filter is already in place at
this position, unscrew it carefully and put it back in its cover.
7
Screw the new filter into place.
8
Select the appropriate filter from the drop-down list and click the OK button.
Note:
The filter changer returns to the original position when you click the Start, Focusing or
Method/Tray position buttons.
4.2.3 Placing the sample
This section describes how to choose an appropriate sample tray, and how to place the sample for
exposure.
1
Select a sample tray suitable for the type of exposure to be performed.
TraySample typeDetection
Epi trayMembraneChemiluminescence
Bioluminescence
NP trayTiter plate
UV trans trayGel (UV Trans illumination)Fluorescence
Epi trayGel (Epi illumination)
Epi trayMembrane