10.
Cables are included in a long plastic bag to limit the cables crossed. This plastic bag can
be kept on site in the gullet, depending on local regulations.
Illustration 6-30:
11.
Pull out the cable set from the spool. A specific pull wire
is attached to the opposite side
of the cable set. This pull wire could be used to pull the full cable set inside the raceway, gullets
or other to facilitate the run in hospital.
Illustration 6-31:
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
218 18 PIST0017 - System Cable Routing