IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
4 - 8 Section 4-2 - General procedures
4-2-3 Power shut down (cont’d)
• Sleep Mode R2.x.x and R3.x
Use this button to reduce the boot up time (is less than 120 seconds). Intended to help on portable
studies.
Sleep saves all Memory content, in the Hard drive and shuts down. When booting up, memory
content is restored from HDD. This feature is not a substitute of a regular shutdown. The system
should be shutdown completely at least once a day to prevent performance issues.
Figure 4-5 Sleep Mode