GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
8 - 192 Section 8-8 - Casters and Brakes replacement
Section 8-8
Casters and Brakes replacement
8-8-1 Purpose of this section
This section describes how to replace the replaceable parts for the Casters and Brakes.
8-8-2 Rear Casters replacement
8-8-2-1 Manpower
Two people, 15 minutes.
8-8-2-2 Tools
For tools needed, refer to: 8-2-5 "Tools needed for servicing the LOGIQ E9" on page 8-5.
NOTE: This procedure requires an extended HEX key (UNBRAKO / ALLEN key) due to the high torque:
130 Nm {95.9 lbf-ft}.
8-8-2-3 Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system
2.) Disconnect the mains power cable from the wall outlet.
3.) Disconnect all Probes and External I/O Cabling.
4.) Remove both Side Covers.
5.) Remove the Rear Bumper.
6.) Remove the Rear Cover.
Follow these links if you need more information:
4-2-3 "Power shut down" on page 4-7.
8-5-2-1 "Side Covers removal" on page 8-31.
8-5-14 "Rear Bumper replacement" on page 8-64.
8-5-11 "Rear Cover replacement" on page 8-54.