IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
9 - 84 Section 9-17 - Internal Cables
Section 9-17
Internal Cables
9-17-1 Top Console Cables
Figure 9-3 Top Console cable diagram - R3.x and earlier
OP PANEL
REG
5V12V
3.3V
CONTROLLER BOARD
BULKHEAD
BOARD
48V
USB
USB
AUDIO
uP
3.3V
KEYBOARD
USB
HUB
5V
USB
ON/
OFF
10.4"
LCD
+12V
HV INV
USB
UP/DN/LOCK
BUTTONS
TRACKBALL
SWITCH
BOARD
+12V
+5V
GEL WARMER
TEMP
CONTROL
MAIN LCD
DISPLAY
BULKHEAD
PLATE
DVI VIDEO
48V POWER/AUDIO
USB FOR OP PANEL
USB FOR VIDEO ON TOUCH PANEL
UP/DN/BRAKE/RELEASE SWITCHES
MEMORY STICK
MICROPHONE
USB/
VIDEO
TOUCH
E
G
U
W
W
+48V
5V
P7
P8
P1
P2
P3
P4
J24
J22
J21
J25
J23
J18
J19
J15
J1
P5
P6
L
5V FROM J21
5V FROM J22
5V
+12V
USB
HUB
USB/
AUDIO
MIC
5V
48V
5V
REG
12V
USB
WLAN
VIDEO
CNTL
F
HV INV
FLAT
SCREEN
R
TO/FROM
MAIN
CONSOLE
CABLE: D
R2.x.x and R3.x.x
OPTION

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor