IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
4 - 36 Section 4-3 - Functional checks
4-3-7 System CFM and PWD Checks
4-3-7-1 Introduction
For information on the system’s different modes as well as expected results, the Basic User Manual or
User Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
For complete information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter
5.
4-3-7-2 Preparations
Use a phantom (optional) when doing these tests.
1.) Connect one of the probes, to the scanner’s active probe connector.
- see: Section 3-6-6 "Connecting Probes" on page 3-20 for info about connecting the probes.
- For available probes, see: Section 9-18 "Probes" on page 9-104.
2.) Turn ON the scanner.
4-3-7-3 Color Mode Checks
Color Flow screens are B or M-Mode screens with colors representing blood or tissue movement. Color
Flow may be selected both from B-Mode, or from M mode, or a combination of these.
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-4 Color B-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-7-5 Color M-Mode
For information, refer to the appropriate version of the LOGIQ E9 Basic User Manual, Chapter 5.
• Check Horizontal Sweep, PRF, Baseline, and Invert
• Check Variance, Color Maps and Cineloop
• Check Sample Volume
• Check Frequency and Acoustic Power
•Adjust ROI